Error Messages 5
Keysight E3633A and E3634A User’s Guide 167
Calibration Errors
The following errors indicate failures that may occur during a calibration. Refer to
the Service Guide for more information.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the power
supply. When applicable, this error will occur at power-on to warn you that the
power supply is unsecured.
702
Cal secured
The power supply is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to unsecure or
secure the power supply. You must use the same security code to unsecure the
power supply as was used to secure it, and vice versa. The security code may
contain up to12 alphanumeric characters. The first character must be a letter.
704
Secure code too long
A security code was received which contained more than 12 characters.
705
Cal aborted
A calibration in progress is aborted when you press any front-panel key, send a
device clear, or change the local/remote state of the instrument.
708
Cal output disabled
Calibration is aborted by sending
OUTP OFF
command during calibrating a
output.