5 Error Messages
168 Keysight E3633A and E3634A User’s Guide
712
Bad DAC cal data
The specified DAC calibration values (
CAL:VOLT
or
CAL:CURR
) are out of range.
Note that the new calibration constants are not stored in the non-volatile
memory.
713
Bad readback cal data
The specified readback calibration values (
CAL:VOLT
or
CAL:CURR
) are out of
range. Note that the new calibration constants are not stored in the non-volatile
memory.
714
Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
715
Bad OCP cal data
The overcurrent protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
716
Bad DAC DNL error correction data
Invalid data measured during the calibration for DAC differential nonlinearity
error correction.
717
Cal OVP or OCP status enabled
Overvoltage protection status or overcurrent protection status is enabled. You
must set both overvoltage and overcurrent protection status to OFF before and
during the calibration.
718
Gain out of range for Gain Error Correction
The slope of the DAC gain is out of range. Hardware fails.
740
Cal checksum failed, secure state

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E3634A and is the answer not in the manual?

Keysight E3634A Specifications

General IconGeneral
Power Rating100 W
Number of Outputs3
Load Regulation≤ 0.01% + 2 mV
Storage Temperature Range-40 to 70 °C
Ripple and Noise≤ 350 μVrms / 2 mVpp
Line Regulation0.01% + 2 mV
Programming Accuracy (Current)0.2% + 5 mA
Operating Temperature Range0 to 40 °C

Related product manuals