Error Messages 5
Keysight E3633A and E3634A User’s Guide 157
Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have inserted a
character such as #, $, or % in the command keyword or within a parameter.
Example:
OUTP:STAT #ON
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a blank
space before or after a colon in the command header, or before a comma.
Example:
VOLT:LEV ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used a
comma instead of a colon, semicolon, or blank space - or you may have used a
blank space instead of a comma.
Example:
TRIG:SOUR,BUS or APPL 1.0 1.0
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may have
entered an extra parameter, or you added a parameter to a command that does
not accept a parameter.
Example:
APPL? 10