126 Keysight E3633A and E3634A User’s Guide
*RST
This command resets the power supply to its power-on state as follows
Command E3633A state E3634A State
CURR 20 A 7 A
CURR:STEP 0.32 mA (typical value) 0.13 mA (typical value)
CURR: TRIG 20 A 7 A
CURR: PROT 22.0 A 7.5 A
CURR:PROT:STAT ON ON
DISP ON ON
OUTP OFF OFF
OUTP:REL OFF OFF
TRIG:DEL 0 0
TRIG:SOUR BUS BUS
VOLT 0 V 0 V
VOLT:STEP 0.36 mV (typical value) 0.95 mV (typical value)
VOLT:TRIG 0 V 0V
VOLT:PROT 22.0 V 55.0 V
VOLT:PROT:STAT ON ON
VOLT:RANG P8V (Low) P25V (Low)