CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unsecured and the output 
state is ON. It enters a voltage value that you obtained by reading an external 
meter. You must first select the minimum calibration level (
CAL:VOLT:LEV MIN
) 
for the value being entered. You must then select the middle and maximum 
calibration levels (
CAL:VOLT:LEV MID
 and 
CAL:VOLT:LEV MAX
) for the value 
being entered. Three successive values must be selected and entered. The 
power supply then computes new voltage calibration constants. These 
constants are then stored in non-volatile memory.
CALibration:VOLTage:LEVel {MINimum|MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output 
state is ON. It sets the power supply to a calibration point that is entered with 
CAL:VOLT
 command. During calibration, three points must be entered and the 
low-end point (MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of the power 
supply. It takes about 10 seconds to execute the command. The calibration 
must be unsecured and the output be opened before calibrating the 
overvoltage protection circuit. The power supply automatically performs the 
calibration and stores the new overvoltage constant in nonvolatile memory. 
Notice that voltage calibration precedes before sending this command.