Error Messages 5
Keysight E3633A and E3634A User’s Guide 163
-430
Query DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but all
data is lost.
-440
Query UNTERMINATED after indefinite response
The
*IDN?
command must be the last query command within a command string.
Example:
*IDN?;:SYST:VERS?
501
Isolator UART framing error
502
Isolator UART overrun error
503
[a]
SPI data error
Data error was detected during the communication between the main controller
U101 and the I/O controller U802.
511
RS-232 framing error
512
RS-232 overrun error
513
RS-232 parity error
514
Command allowed only with RS-232
There are three commands which are only allowed with the RS-232 interface:
SYSTem:LOCal
,
SYSTem:REMote
, and
SYSTem:RWLock
.
521
Input buffer overflow
522
Output buffer overflow
[a] This error message is only applicable for serial MY53xx6xxx.