3 Front-Panel Operation
60 Keysight E3633A and E3634A User’s Guide
– Adjust the output voltage level
1 Lower the output voltage level.
Lower the output voltage level below the OVP trip point after pressing   
key. The OVP and Limit annunciators are blinking.
2 Move to the clear mode.
Press   key twice to move to the OVP CLEAR mode. The “OVP ON” 
message appears on the display. Turn the knob to the right until the above 
message appears on the display.
3 Clear the overvoltage cond ition and exit this menu.
Now, when you press   key again, the “DONE” message is displayed for 
a second and the OVP annunciator will not blink any more. The output will 
return to meter mode.
OVP CLEAR