Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide 45
M8194A Soft Front Panel 3
Single Channel with Markers
If this mode is selected, Channel 1 is used to generate data, and
channel 3 and 4 are used to generate digital markers. Channel 2 is
disabled.
Dual Channel
If this mode is selected, Channel 1 and Channel 2 are used to
generate data. Channel 3 and Channel 4 are disabled. This mode is
selectable, if option 002 or 004 is present.
Dual Channel with Markers
If this mode is selected, Channels 1 and 2 are used to generate a
signal. Channel 1 has two markers output on channel 3 and 4.
Channel 2 can generate a signal without markers.
Four Channel
This mode is only selectable, if option 004 is installed. If this mode is
selected, all four channels can be used to generate data.
Channel Settings
Each channel in the Output tab has the following input fields:
Amplitude - Sets the single-ended, peak-to-peak amplitude of the
output signal.
Offset - Sets the common mode offset voltage of the output signal.
Diff. Offset (Differential Offset) - Sets the differential offset of the
output signal.
VTerm (Termination Voltage) - Sets the termination voltage.
Output enable check box
When this check box is selected, it enables the output so that the
generated signal is present at the output.
Output status indicator. This indicator reflects the color of the
‘Channel’ LED on the front panel:
It is ‘OFF’ when the channel is disabled and no overload
condition at this channel has been detected.
It is ‘GREEN’ if the channel is enabled and no overload
condition at this channel has been detected.

Table of Contents