Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide 45
M8194A Soft Front Panel 3
• Single Channel with Markers
If this mode is selected, Channel 1 is used to generate data, and
channel 3 and 4 are used to generate digital markers. Channel 2 is
disabled.
• Dual Channel
If this mode is selected, Channel 1 and Channel 2 are used to
generate data. Channel 3 and Channel 4 are disabled. This mode is
selectable, if option 002 or 004 is present.
• Dual Channel with Markers
If this mode is selected, Channels 1 and 2 are used to generate a
signal. Channel 1 has two markers output on channel 3 and 4.
Channel 2 can generate a signal without markers.
• Four Channel
This mode is only selectable, if option 004 is installed. If this mode is
selected, all four channels can be used to generate data.
• Channel Settings
Each channel in the Output tab has the following input fields:
• Amplitude - Sets the single-ended, peak-to-peak amplitude of the
output signal.
• Offset - Sets the common mode offset voltage of the output signal.
• Diff. Offset (Differential Offset) - Sets the differential offset of the
output signal.
• VTerm (Termination Voltage) - Sets the termination voltage.
• Output enable check box
When this check box is selected, it enables the output so that the
generated signal is present at the output.
• Output status indicator. This indicator reflects the color of the
‘Channel’ LED on the front panel:
• It is ‘OFF’ when the channel is disabled and no overload
condition at this channel has been detected.
• It is ‘GREEN’ if the channel is enabled and no overload
condition at this channel has been detected.