88 Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide
3 M8194A Soft Front Panel
Waveform Destination Section
Channel
Independent check boxes allow to import waveforms for Channel 1,
Channel 2, Channel 3 or Channel 4. One of the boxes is always
selected. If the file contains only one waveform, when pressing the
‘Send To Instrument’ button, the waveform is sent to all channels that
are selected. If the file contains multiple waveforms (file types
MAT89600 and CSV), they can be sent to multiple channels in one
operation.
The following two tables show the standard column-to-channel
mapping for the case of no additional data header in the CSV file or no
reordering of the column names in the MAT89600 file.
Table 8 Standard column-to-channel mapping in four-channel mode
Table 9 Standard Column to channel mapping in two-channel mode
Number of columns in
file for real values
Import and download to M8194A, when corresponding channel
box is selected
1. Column 1 to Ch 1 and Ch 2 and Ch 3 and Ch4
2. Column 1 to Ch 1 and Column 2 to Ch 2
3. Column 1 to Ch 1 and Column 2 to Ch 2 and Column 3 to Ch 3
4. Column 1 to Ch 1 and Column 2 to Ch 2 and Column 3 to Ch 3 and
Column 4 to Ch 4
Number of columns in
file for real values
Import and download to M8194A, when corresponding channel
box is selected
1. Column 1 to Ch 1 and Ch4
2. Column 1 to Ch 1 and Column 2 to Ch 4
3. Column 1 to Ch 1 and Column 2 to Ch 4, Column 3 is ignored
4. Column 1 to Ch 1 and Column 2 to Ch 4, Column 3 and 4 are
ignored

Table of Contents