Electrical Requirements
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 4–3
Electrical Requirements
WARNING*
You MUST select the operating voltage when placing an order. The
operating voltage CANNOT be modified on site.
The unit can operate at:
100/110/130 V 50/60 Hz
230/240 V 50/60 Hz
CAUTION
The power supply line must be equipped with a connection box that
ensures a constant connection. It must not be possible to connect the
unit to the power supply without using a tool. The unit must be
protected against any accidental disconnection.
If other units are installed on the same line, interference and voltage fluctuations can
cause the radiological unit to operate abnormally. We strongly recommend that a
separate electrical line be dedicated to supply power to the KODAK 8000C unit.
This line should be protected by a circuit breaker with a maximum current of:
16 A at 230/240 V
20 A at 110/130 V
A differential circuit breaker of 30 mA
Table 4–1 Optional Operating Voltages of the Unit
Nominal voltage *
(no load)
Minimum Maximum
Maximum line
current
100 V - 130 V 90 V 143 V 20 A
230 V - 240 V 207 V 264 V 10 A