Post-Installation Control
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02 5–17
Adjusting the Cephalometric Frankfurt Positioning Laser
To adjust the cephalometric Frankfurt positioning laser, follow these steps:
1. Remove the generator front cover.
2. Turn on the unit. On the control panel, press to turn on the positioning laser
beams.
3. Make sure that the cephalometric Frankfurt positioning laser beam is projected in the
middle of the ear cone.
4. If needed, adjust the laser using the fixing screw of the laser socket.
5. Replace the generator front cover.