Post-Installation Control
5–14 INSTALLING THE CEPHALOSTAT
Adjusting the Ear Cones Alignment
To adjust the ear cones alignment, follow these steps:
1. Position the head clamps in a lateral position.
2. Access the Cephalometric Acquisition Window. In the Program pane, click .
3. Launch an x-ray using the x-ray remote control. The ear cone ball (A) must be at the
center of the head clamps circle (B).
4. In the Imaging Window, on the drawing toolbar, click . On the acquired image,
draw a line from the center of the ball (A) to the center of the circle (B). The resulting
value must be
±1.5mm.
A
B
+1.5 mm