Electronic Copy of LTI’s UltraLyte User’s Manual – 7
th
Edition June 1998
43
Power Conservation Intervals
To help conserve its batteries, the UltraLyte has
three timed power-off intervals: one for the in-scope
numerical display, one for the in-scope aiming dot,
and one for the instrument as a whole. Lack of
instrument activity for an interval causes the
associated function to shut down. The following
table summarizes the time-outs:
Function Interval Effect of Time-out
In-scope
numerical display
3 sec. The numerical display
shuts off
In-scope aiming
dot
20 sec. Aiming dot shuts off
Instrument 10 min.
(approx.)
Power shuts down
The in-scope display time-outs cannot be turned off.
The instrument time-out, however, can be.
Follow this procedure:
Action Result
1. Press Menu/Options
repeatedly until the
ALL on
indicator
blinks, and the screen
shows On (or OFF).
MENU
2. Press the Select/Edit
button to turn the time-
out on or off.
MENU
3. Return to speed or survey mode by pressing the
appropriate button. Go to the next menu option by
pressing Menu/Options.
Use care.
Disabling the power-off interval prevents
the instrument from automatically shutting off; it must
be manually turned off or the battery life will be
greatly reduced.