Personnel:
n
Operating personnel
CAUTION!
Bursting of external consuming unit due to excess pressure
Scalding, frostbite, cuts
If using a pressure-sensitive consuming unit (such as a glass
reactor), use a pressure relief device.
CAUTION!
Bursting of the external application due to overpressure
Scalding, frostbite, impacts
For applications with a maximum permissible working pres-
sure that is less than the maximum pressure of the pump,
use a pressure relief device for protection. This pressure
relief device must be installed in the outflow to the applica-
tion.
Set the maximum pump pressure as per your application.
1. To increase the pressure in the consuming unit, turn the bypass
adjusting wheel clockwise until the maximum permissible pressure for
the external consuming unit is reached.
Monitor the pressure indication on the front of the device
when doing this.
V08 Variocool 35 / 97