Table 16: Power consumption of VC 1200 (W) to VC 3000 (W)
Alternating current VC 1200 (W) VC 2000 (W) VC 3000 (W) Unit
230 V; 50 Hz 1.1 1.6 1.8 kW
200 V; 50/60 Hz 1.3 2.0 2.2 kW
208-220 V; 60 Hz 1.4 2.2 2.3 kW
Table 17: Power consumption of VC 5000 (W) to VC 10000 (W)
Three-phase current VC 5000 (W) VC 7000 (W) VC 10000 (W) Unit
400 V; 3/N/PE~50 Hz 3.3 4.3 5.4 kW
208-220 V; 3/PE~60 Hz 3.6 4.6 5.9 kW
200 V; 3/PE~50/60 Hz 3.5 4.5 5.7 kW
11.2 Cooling capacity
Table 18
Device
Cooling capacity
(20 °C)
Cooling capacity
(10 °C)
Cooling capacity
(0 °C)
Cooling capacity
(-10 °C)
Cooling capacity
(-20 °C)
kW kW kW kW kW
VC 1200 (W) 1.20 1.00 0.70 0.40 0.14
VC 2000 (W) 2.00 1.50 1.06 0.68 0.38
VC 3000 (W) 3.00 2.40 1.68 0.95 0.45
VC 5000 (W) 5.00 3.90 2.75 1.70 0.90
VC 7000 (W) 7.00 5.30 3.70 2.40 1.30
VC 10000 (W) 10.00 7.60 5.30 3.50 2.00
The cooling output is measured when the heat transfer liquid rea-
ches a certain temperature. These temperature values are specified
in brackets. The ambient temperature for the measurement is 20 °C
and ethanol was used as a heat transfer liquid. To measure water-
cooled devices, the cooling water temperature is 15 °C and the
cooling water dierential pressure is 3 bar.
All water-cooled Variocool circulation chillers have the following cooling
water connection:
n ¾" external connection thread
Power consumption without a heater
Cooling water connection