Personnel:
n
Operating personnel
1. Press and hold down the [ENTER] button.
2. Press and hold down the [Up] arrow button within 4 seconds.
3. Hold down both buttons for 4 seconds.
In the display the descriptions of the buttons show up again.
The device can be operated again.
6.6 Define temperature limits
You can use the temperature limits to specify the temperature range of your
application, i.e. the temperature range in which temperature control can take
place.
Personnel:
n
Operating personnel
1. Switch to the main menu.
2.
Select the menu item Setup Temp. limits.
3. Select one of the following options:
n Select the first entry Til to set the lower limit value.
n Select the second entry Tih
to set the upper limit value.
4. Customize the value in the following entry window.
Release operating buttons
Fig. 23: Selecting the temperature limit
Fig. 24: Defining temperature limits
V08 Variocool 43 / 97