8.5 Warnings - control system
All warnings of the control system start with the prefix 0. The prefix
is followed by two additional numbers. These numbers are listed in
the following table.
Code English output Description
001 CAN receive overflow Overflow during CAN reception
002 Watchdog reset Watchdog reset
003 T_il limit active til-limit active
004 T_ih limit active tih-limit active
005 Corrupt parameter Inadmissible internal parameter
006 Corrupt programme Inadmissible programmer data
007 Invalid parameter Inadmissible parameter in memory
008 CAN system Problem during internal data exchange
009 Unknown module Unknown module connected
010 SW Control too old Software version of control system too old
011 SW Safety too old Software version of safety system too old
012 SW Command too old Software version of remote control unit Command too old
013 SW Cool too old Software version of cooling module too old
014 SW Analogue too old Software version of analogue module too old
015 SW Serial too old Software version of serial interface (RS 232) too old
016 SW Contact old Software version of contact module too old
017 SW Valve 0 old Software version of solenoid valve 0 too old
018 SW Valve 1 old Software version of solenoid valve 1 too old
019 SW Valve 2 old Software version of solenoid valve 2 too old
020 SW Valve 3 old Software version of solenoid valve 3 too old
021 SW Valve 4 old Software version of solenoid valve 4 too old
022 SW Pump 0 old Software version of pump 0 too old
023 SW Pump 1 old Software version of pump 1 too old
024 SW Pump 2 old Software version of pump 2 too old
025 SW Pump 3 old Software version of pump 3 too old
026 SW HTC old Software version of high temperature cooler too old
027 SW Ext. Pt100 old Software version of external Pt100 too old
028 SW Ethernet old Software version of Ethernet too old