Code English output Description
315 SW Serial too old Software version of serial interface (RS 232) too old
316 SW Contact old Software version of contact module too old
317 SW Valve 0 old Software version of solenoid valve 0 too old
318 SW Valve 1 old Software version of solenoid valve 1 too old
319 SW Valve 2 old Software version of solenoid valve 2 too old
320 SW Valve 3 old Software version of solenoid valve 3 too old
321 SW Valve 4 old Software version of solenoid valve 4 too old
322 SW Pump 0 old Software version of pump 0 too old
323 SW Pump 1 old Software version of pump 1 too old
324 SW Pump 2 old Software version of pump 2 too old
325 SW Pump 3 old Software version of pump 3 too old
326 SW HTC old Software version of high temperature cooler too old
327 SW Ext. Pt100 old Software version of external Pt100 too old
328 SW Ethernet old Software version of Ethernet module too old
329 SW EtherCAT old Software version of EtherCAT too old
341 sm0 min too small Starting value of injection valve too low
344 chiller missing Refrigerant unit does not run
345 Valve not closed Valve of the cooling circuit does not close
347 Configure EEV0 Contact LAUDA Service
348 Configure EEV1 Contact LAUDA Service
349 Preheat unit A warning concerning possible damage to the cooling system is
sent when the device is operated below 5 °C. The cooling system
must be preheated to prevent damage. This is automatically per-
formed when outside installation is activated.
355 CAN bu. overflow Buer overflow for CAN reception
V08Variocool78 / 97