Type l/2 dip
ole antenna
Connector Special customized SMB
7.10.3 Dangerous Goods Regulations
Many products of Leica Geosystems are powered by Lithium batteries.
Lithium ba
tteries can be dangerous under certain conditions and can pose a
safety hazard. In certain conditions, Lithium batteries can overheat and ignite.
☞
When carrying or shipping your Leica product with Lithium batteries
onboard a commercial aircraft, you must do so in accordance with the
IATA Dangerous Goods Regulations.
☞
Leica Geosystems has developed Guidelines on “How to carry Leica
products” and “How to ship Leica products” with Lithium batteries.
Before any transportation of a Leica product, we ask you to consult
these guidelines on our web page
(http://www.leica-geosystems.com/dgr) to ensure that you are in
accordance with the IATA Dangerous Goods Regulations and that the
Leica products can be transported correctly.
☞
Damaged or defective batteries are prohibited from being carried or
transported onboard any aircraft. Therefore, ensure that the condition
of any battery is safe for transportation.
7.11 General Technical Data of the Product
Several factors can influence the system's accuracy for determining the location
o
f a prism:
•
Internal ATR accuracy
•
Angular accuracy of the instrument
•
Type and centring accuracy of the prism
•
Selected EDM measuring program
•
External measuring conditions
Therefore, the overall pointing accuracy of the determined point location can
be lower than the given angular accuracy and the ATR accuracy.
The following paragraphs provide a short overview of these influencing factors
and their possible intensities.
Angular accuracy
The accuracy of angular measurements depends on the instrument type. The
angular accuracy for total stations is typically in the range from 0.5² to 5². The
resulting error depends on the measurement distance.
Angular accuracy
Possible deviation
*
a
t 100 m distance
1² ~0.5 mm
3² ~1.5 mm
*
Orthogonal to the line of sight.
☞
Refer to the data sheet of the respective instrument model for infor-
ma
tion about the angular accuracy.
EDM accuracy
T
he distance measurement accuracy consists of two parts: a fixed value and a
distance-dependent value (ppm-value).
Dangerous Goods
R
egulations
System Accuracy
68 Technical Data