EasyManua.ls Logo

Leuze electronic BCL 8 S 2 Series - 10.1.1 Properties Window - Decode Tab; Figure 10.2: Standard Settings for the Properties Window - Decode Tab

Leuze electronic BCL 8 S 2 Series
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Important Parameters
Leuze electronic BCL 8 47
TNT 35/7-24V
10.1.1 Properties window – Decode tab
Figure 10.2: Standard settings for the Properties window – Decode tab
Quiet zone size
Quiet zone: the area to the left and right of the bar code
Module: width of the narrowest line in the bar code
According to the code specifications, each bar code must have a
quiet zone which is 10 times as wide as the module of the bar
code.
Ex: for a code having a module of 0.5mm, 5mm blank space must
be present at both the left and right of the code.
By default, the scanner checks a quiet zone which is 7 times
greater than the module. This means 7x or greater is acceptable
for the scanner.
Reading security
(equal scans)
Specifies how often a code must be decoded before the result is
valid and output. This value should only be increased for test pur-
poses or for codes with low security.
No time correlation
between two identical
scans
If this parameter is set, a gap between two identical labels is
ignored and they are treated as a single label.
No position
correlation between
two identical scans
If this parameter is set, then the position of a bar code label in the
reading beam is not taken into account. Identical labels are treated
as a single label.

Table of Contents

Related product manuals