EasyManua.ls Logo

logos biosystems LUNA-FX7 Basic - Software Update Procedures

Default Icon
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
Calibration
The LUNA-FX7™ is calibrated prior to shipping. Calibration only needs to be
performed 1) after any software/firmware updates, and 2) after switching trypan blue
or erythrosin B brands or formulations.
To perform calibration:
1. Press DATE & CALIBRATION.
2. Press Start calibration. The Calibration Step 1 window will appear.
3. Remove any counting slide from slide port.
4. Press START.
5. The Calibration Step 2 window will appear.
6. Mix one part stain with an equal volume of distilled water, PBS, or plain medium.
Put 10 μL of the diluted stain into Chamber A of a PhotonSlide.
7. Insert the slide face up into the port, so that Chamber A is to the left.
8. Press START; or, press SKIP if only fluorescence calibration is needed.
9. The Calibration Step 3 window will appear upon completion.

Table of Contents

Related product manuals