Menu Parameter Values Meaning
Set Corr: Adjustment of the measurements **
Zero correction/offset of the temperature
measurement
**
oFF No zero correction of the temperature
measurement
-5.0 -
5.0 %
Zero correction of the temperature
measurement in °C
Slope adjustment of the temperature measurement **
oFF No pitch correction of the temperature
measurement
-5.0 -
5.0 %
Pitch correction of the temperature
measurement in %
Set Alarm: Adjustment of the alarm function
On Measurement channel cond/rES/TDS/
SAL: Alarm on with sound
No.So Measurement channel cond/rES/TDS/
SAL: Alarm on without sound
OFF No alarm function for measurement
channel cond/rES/TDS/SAL
0.0000 µS/cm
... 1000 mS/cm
Min. alarm limit cond/rES/TDS/SAL (not
with AL. 1. oFF)
0.0000 µS/cm
... 1000 mS/cm
Max. alarm limit cond/rES/TDS/SAL (not
with AL. 1. oFF)
On Alarm temperature measurement on
with sound
No.So Alarm temperature measurement alarm
on without sound
OFF No alarm function for temperature
measurement
-5.0 -
+100.0 °C
Temperature min. alarm threshold (not
with Al. 2. oFF)
-5.0 -
+100.0 °C
Temperature max. alarm threshold (not
with Al. 2. oFF)
(**) If logger is running parameters marked with (**) cannot be called.