Setpoints
BPD17 pH / Redox Monitor
Operating Guide
- 47 -
Action
Set the setpoint to work in the normal mode or
reverse mode - which is akin to a normally closed
relay except it will fall open if the power to the
instrument is removed.
/ – Select Option
EXIT
– Cancel
– Save Selection
Low Value
The Setpoint Low value.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
High Value
The Setpoint High value.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Delay
In order to prevent short duration changes at the
input affecting the setpoint operation a delay can be
set before the setpoint is energised. If the input is
still the same after the delay, then the setpoint will be
energised.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value