Setpoints
BPD17 pH / Redox Monitor
Operating Guide
- 51 -
Setpoint Cleaning Mode
The Setpoints can be configured to operate a jet spray wash or rotary electrode cleaning system on a
timed cycle. Its purpose is to prevent accumulation of particulate matter on the active surfaces of the
sensor. Note when cleaning is active the input will be taken offline, this will prevent any undesired
control actions resulting from spraying cleaning solution onto the sensor.
Trigger
If available select cleaning from the list of options.
/ – Select Option
EXIT
– Cancel
– Save Selection
Clean Duration
Enter the duration of the cleaning operation.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Clean Interval
Enter the time between cleaning operations.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value