EasyManua.ls Logo

Magnamed Fleximag - Page 97

Magnamed Fleximag
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96
MAGNAMED TECNOLOGIA MÉDICA S/A
Medium Priority Alarm
Delay time
Description
LOW PEEP
< 3 cycles
Positive pressure end-expiratory
pressure (PEEP) did not reach the alarm
value set to its lower limit.
HIGH FIO
2
< 3 cycles
The fraction of inspired O
2
exceeded the
alarm value set to its upper limit.
LOW FIO
2
< 3 cycles
O
2
fraction of inspired not hit the alarm
value set to its lower limit.
LIMITED PRESSURE
< 1 s
When the monitored pressure reaches
the set pressure limit. In this case the
volume delivered by the ventilator
module NOT REACHED the adjusted
volume due to pressure limitation.
Upon receiving alarm information, provide prompt service to solve the problem.
So, it is terminated the situation that necessitated the total audible alarm silence, one should re-
enable it for patient safety.
To silence the audible alarm, press the quick access button SILENCE. The audible alarms are
disabled for the set period or until a new alarm occurs.
There may be dangerous if different pre-alarm settings used for the same equipment or similar
equipment in the same area, for example, an intensive care unit or operating room heart.
The equipment always starts with the volume set audio to the maximum level (6), regardless of
the level set when it was switched off.
If the volume of audio is set to a value below the maximum level (6), if an alarm while there is no
answer to this cease, the volume of audio will be increased gradually every 15 seconds to reach
its maximum.

Table of Contents

Related product manuals