5 Annex
5.1 Maintenance and care
Make sure that the measuring plate is always
kept clean. Dust, oil and coolant emissions
must be removed from the measuring plate
every day.
Dirt on the air bearings has a detrimental effect
on measuring behavior and accuracy.
The battery can be changed without the data
stored in the instrument being lost.
5.2 Cleaning the instrument
The instrument can be cleaned using a damp
cloth. When cleaning, do not allow any liquids
to penetrate the instrument!
Do not use solvents or detergents that are
corrosive or harmful to plastics for cleaning.
The air bearings can be cleaned with
methylated spirits (alcohol).
5.3 Charging the battery
To charge the battery, the AC adapter must be
connected to the charging socket. The charge
status of the battery is displayed in the status
bar of the user interface.
The AC adapter can also be permanently
connected, as an overload fuse monitors the
charging process. The battery continues to
charge when the height measuring instrument
is switched off.
Batteries discharge over time even
when they are not in use. Discharged
batteries can lose their capacity or
become completely unusable. The
battery should therefore be charged
every 6 months.
Never short-circuit the battery. There
is a risk of fire and explosion!
The height measuring instrument
should always be operated with the
battery inserted.