7.0 Specifications
MODELS: LONG RANGE SHORT RANGE
Digital Range:
Standard Range
r0.100 in / r199 mm r0.040 in / r100 mm
Reduced Range None
r0.0199 in / r0199 mm
Digital Resolution
0.0001 in / 0001 mm 0.00002 in / 00005 mm
(switchable^)
0.0005 in / 0005 mm 0.00005 in / 0001 mm
0.0001 in / 0001 mm
0.0005 in / 0005 mm
Analog Range:
25 or 50 5 or 25 or 50
Analog Resolution:
0.001 in / 001mm 0.001 in / 001 mm
(switchable^)
0.0005 in / 0005mm 0.0005 in / 0005 mm
0.0001 in / 0001 mm
Total Spindle Travel:
0.250 in / 635mm 0.120 in / 03 mm
Pre-Travel^:
0.020 in / 05mm 0.010 in / 03 mm
Over-Travel^:
0.030 in / 08mm 0.030 in / 0,8 mm
Accuracy:
Standard Range 0.35% of Total Range 0.5% of Total Range
Reduced Range N/A 0.25% of Total Range
Repeatability:
r One Least Digit r One Least Digit
Gaging Force:
3-4oz./85-112 gM, preloaded
Operation Temperature:
50q to 130q F / 10q to 55qC
Storage Temperature:
-4q to 149qF / -20q to 65qC
IP Rating:
54
Weight:
!pproximately 13.6 oz / 385 gm
Battery Life:
!pproximately 4000 hours
Operating Voltage 2.8 to 3.1VDC
^ - $ependent on standard or reduced range selected. Standard Range shown.