Masimo LiDCO Chapter 4: Operation
www.masimo.com 34 Masimo
Options
Description
Factory
Default
Selectable Settings
Display
parameter
how parameter is displayed on the
monitor
Absolute
Absolute, Index or
Both
CO Calibration
Options
Description
Factory
Default
Selectable
Settings
CO(L/min)
reflects the patient-specific maximum aortic
capacitance and will generally remain constant over
short periods of time.
NA
NA
Measure
CO(L/min)*
known value for Cardiac Output (CO) taken during
hemodynamically stable period with minimal variation
in blood pressure or heart rate
Off
Off, or .1-30 in
steps of 0.1
* Calibration can only be adjusted during an active session.
Cardiac Index (Ci)
Alarms
Cardiac Index (Ci) will display only when Index or Both is selected for Cardiac Output parameter display.
Options
Description
Alarm
Priority
Faculty
Default
Selectable
Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 0.3-14.9 in
steps of 0.1
Low Limit
Lower limit that triggers an alarm
Low/Medium*
Off
Off, or 0.2-14.8 in
steps of 0.1
High Caution
Range
Value below High Limit to trigger
caution light
NA
Off
Off, or 0.1-1.1 in
steps of 0.1
Low Caution
Range
Value above Low Limit to trigger
caution light
NA
Off
Off, or 0.1-1.1 in
steps of 0.1
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30s, 1m, 2m, 5m.
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest Ci value that can be displayed
5
0.1-15 in steps of 0.1
Y-Axis Min
Lowest Ci value that can be displayed
.5
0.0-14.9 in steps of 0.1