Masimo LiDCO Chapter 4: Operation
www.masimo.com 42 Masimo
Systemic Vascular Resistance Index (SVRi)
Systemic Vascular Resistance Index (SVRi) will display only when Index is selected for Systemic
Vascular Resistance parameter display.
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 850-3000 in
steps of 50,
Low Limit
Lower limit that triggers an alarm
Low/Medium*
Off
Off, or 800-2950 in
steps of 50
High Caution
Range
Value below High Limit to trigger
caution light
NA
Off
Off, or 100-600 in
steps of 50
Low Caution
Range
Value above Low Limit to trigger
caution light
NA
Off
Off, or 100-600 in
steps of 50
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30 sec, 1, 2, or 5
min
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
Trends
Options
Description
Factory
Default
Selectable Settings
Y-Axis
Max
Highest SVRi value that can be
displayed
3000
1000-12000 in steps of
1000
Y-Axis Min
Lowest SVRi value that can be
displayed
0
0 to 11000 in steps of 1000
Guided Protocols
Masimo LiDCO provides Guided Protocols to assist in performing fluid response assessments. They
provide supporting information and guidance to complete each along with a detailed display of the