Masimo LiDCO Chapter 4: Operation
www.masimo.com 43 Masimo
hemodynamic parameters. Touch Guided Protocols for drop-down menu. This menu option is not
available when SVi is being displayed.
Guided protocols(1) are intended to aid the clinician in giving a fluid challenge and observing the effects
on advanced hemodynamic parameters including Stroke Volume (SV), Mean Arterial Pressure (MAP),
Heart Rate (HR) and Stroke Volume Variation (SVV) if applicable.
Use the Additional Settings screen to configure the following:
Ref
Options
Description
Purpose
2
Fluid
Challenge
Giving a small amount of fluid in a short period of time
to assess whether the patient has a preload response
that can be used to increase the stroke volume with
further fluids.
If SV increases by
≥10% then likely fluid
responsive
3
Passive Leg
Raise
Using the bed to position the patient's upper body at
zero degrees while simultaneously raising both legs to
about 45 degrees 30-90 seconds.
If SV increases by
≥10% then likely fluid
responsive
4
End Expiratory
Occlusion Test
Occluding the airway at the end of expiration for 10-30
seconds and observe the maximum rise in SV.
If SV increases by
≥5% then likely fluid
responsive