Chapter 7: Service and Maintenance
Cleaning Procedures
The module for Masimo LiDCO is a reusable instrument. The instrument is supplied non-sterile.
To clean the module:
1. Disconnect the module from the monitor and ensure the cable is not connected to the
module.
2. Wipe the outer surfaces using a dampened soft cloth with a mild detergent and warm water
solution or one of the recommended cleaning solutions twice or until the surfaces are free of
any visible residue.
3. Dry the module thoroughly prior to using on a patient.
The surfaces of the module may be cleaned with the following solvents or cleaning agents:
70% Isopropyl Alcohol (IPA)
Glutaraldehyde (Cidex
®
Plus)
0.5% sodium hypochlorite water solution (10% bleach/water solution)
Quaternary ammonium chloride
Accelerated Hydrogen Peroxide
®
solutions (Oxivir
®
TB)
See Safety Information Warnings and Cautions on page 9
General Maintenance for Module
Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should
be performed at regular intervals or in accordance with hospital, as well as local and governmental
regulations.
The following is a checklist for the general maintenance of the module:
Visually inspect equipment for functional or structural damage, including poor seals, cracks,
damaged springs, etc.
Visually inspect cables, connectors, and connector pins for signs of damage or wear.
Visually inspect product identification labels to ensure they are clear and legible.
Visually inspect for evidence of fluid ingress.
Service Instructions
The module has no customer serviceable parts. Attempting to service the module will void the warranty.
Safety tests and internal adjustments should be done by qualified personnel only.
See Sales & End-User License Agreement on page 61.
See Contacting Masimo on page 60.