Masimo LiDCO Chapter 4: Operation
www.masimo.com 41 Masimo
Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest SVV value that can be displayed
50
5 to 100 in steps of 5
Y-Axis Min
Lowest SVV value that can be displayed
0
0 to 95 in steps of 5
Systemic Vascular Resistance (SVR)
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 450-1500 in
steps of 50.
Low Limit
Lower limit that triggers an alarm
Low/Medium*
Off
Off, or 400-1450 in
steps of 50.
High Caution
Range
Value below High Limit to trigger
caution light
NA
Off
Off, or 50-300 in
steps of 50.
Low Caution
Range
Value above Low Limit to trigger
caution light
NA
Off
Off, or 50-300 in
steps of 50.
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30s, 1m, 2m, 5m.
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest SVR value that can be displayed
1500
100-12000 in steps of 100
Y-Axis Min
Lowest SVR value that can be displayed
300
0-11900 in steps of 100
Additional Settings
Option
Factory Default
Selectable Settings
Display parameter
Absolute
Absolute or Index
CVP
7
1-20 in steps of 1