Masimo LiDCO Chapter 4: Operation
www.masimo.com 40 Masimo
Stroke Volume Index (SVi)
Stroke Volume Index (SVi) will display only when Index is selected for Stroke Volume parameter display.
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 10-245 in
steps of 5
Low Limit
Lower limit that triggers an alarm
Low/Medium*
Off
Off, or 5-240 in
steps of 5
High Caution
Range
Value below High Limit to trigger
caution light
NA
5
Off, or 5-15 in
steps of 5
Low Caution
Range
Value above Low Limit to trigger
caution light
NA
5
Off, or 5-15 in
steps of 5
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30 sec, 1, 2, or 5
min
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off
Trends
Options
Description
Factory Default
Selectable Settings
Y-Axis Max
Highest SVi value that can be displayed
75
5-250 in steps of 5
Y-Axis Min
Lowest SVi value that can be displayed
15
0-245 in steps of 5
Stroke Volume Variation (SVV)
Alarms
Options
Description
Alarm
Priority
Factory
Default
Selectable
Settings
High Limit
Upper limit that triggers an alarm
Low/Medium*
Off
Off, or 2-29 in
steps of 1.
High Caution
Range
Value below High Limit to trigger
caution light
NA
Off
Off, or 1-13 in
steps of 1.
Silence
Duration
Length of time that the audible
alarm remains silenced
NA
2 min
30s, 1m, 2m, 5m.
Audible
Alarms*
Disables audible alarm
NA
Off
On or Off