Page 23 of 54 5100054 BND01O03.docx
10.6. Assembly of the Injector Holder
1. Install the Tablet to the Holding
Plate with 3x countersunk
screws (Allen Key Size 3)
2. Tighten the screws to a torque of 5
Nm.
3. Screw in the component grounding
screw with the spring washer
and the flat washer and tighten
in to a torque of 2 Nm.
Figure 15 Assembly of the Injector Holder
4. Install the Holding Plate to the
Injector Column with 4x
countersunk screws (Allen Key
Size 3).
5. Tighten the screws to a torque of 5
Nm.
6. Screw in the component grounding
screw with the spring washer
and the flat washer and tighten
in to a torque of 2 Nm.