Page 23 of 54 5100054 BND01O03.docx
10.6. Assembly of the Injector Holder
1. Install the Tablet ② to the Holding
Plate ③ with 3x countersunk
screws ① (Allen Key Size 3)
2. Tighten the screws to a torque of 5
Nm.
3. Screw in the component grounding
screw ⑥ with the spring washer ⑤
and the flat washer ④ and tighten
in to a torque of 2 Nm.
Figure 15 Assembly of the Injector Holder
4. Install the Holding Plate ③ to the
Injector Column ⑦ with 4x
countersunk screws ① (Allen Key
Size 3).
5. Tighten the screws to a torque of 5
Nm.
6. Screw in the component grounding
screw ⑥ with the spring washer ⑤
and the flat washer ④ and tighten
in to a torque of 2 Nm.
⑦
③
①
⑥
④
①
②
③
④
⑤