28
starlight pro
11.1.2 AccessiblePartsoftheCasing
The customer or the user of the device should ensure that it is used in such environment.
Phenomenon
BasicEMC
standardortest
method
Immunitytestlevels
Electromagnetic
EnvironmentGuidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV on contact
±15 kV in air
relative humidity should be at
Radiated RF
a)
IEC 61000-4-3
3 V/m
f)
b)
c)
Portable and mobile RF
communication devices
should not be used near any
respect the recommended
equation applicable at the
frequency of the transmitter.
RATED power
frequency
magnetic
d) e)
IEC 61000-4-8
30 A/m
g)
mains frequency should
have levels characteristic
of a typical location in a
commercial or hospital
environment.
a)
b) The device that intentionally receives RF electromagnetic energy for the purpose of its operation shall be tested
when an ambient signal is in the passband. It is understood that the receiver might not achieve normal reception during
the test.
c)
d) Applies only to ME EQUIPMENT and ME SYSTEMS with magnetically sensitive components or circuitry.
e)
signal.
f) Before modulation is applied.
g)
least 15 cm. If the RISK ANALYSIS shows that the device will be used closer than 15 cm to sources of power frequency