29

11.1.3 Guidance and Manufacturers Declaration - Electromagnetic
Immunity
11.1.3.1 PowerConnectionA.C.Input

The customer or the user of the device should ensure that it is used in such environment.
Phenomenon
BasicEMC
standardortest
method
Immunitytestlevels
Electromagnetic
EnvironmentGuidance
Electrical fast
transients /
bursts
a) l) o)
IEC 61000-4-4
±2 kV on contact

frequency
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Surges
a)
b) j) o)
Line-to-line
IEC 61000-4-5 
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Surges
a) b) j) k) o)
Line-to-
ground
IEC 61000-4-5 
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Conducted
disturbances
induced by RF

c) d) o)
IEC 61000-4-6
3 V
m)

6 V
m)
in the ISM bands


n)

e)
Portable and mobile RF
communication devices
should not be used near any


respect the recommended

equation applicable at the
frequency of the transmitter.
Voltage dips
f) p) r)
IEC 61000-4-11




The quality of the network
voltage should be that of a
typical commercial or hospital
environment.


h)

Voltage
interruptions
f) i) o) r)
IEC 61000-4-11 
h)
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
a) The test may be performed at any one power input voltage within the device RATED voltage range. If the device is

b) All the device cables are attached during the test.
c) 
d) 
