32
starlight pro
11.1.3.3 PartsAccessibletotheInput/OutputSignals
The customer or the user of the device should ensure that it is used in such environment.
Phenomenon
BasicEMC
standardortest
method
Immunitytestlevels
Electromagnetic
EnvironmentGuidance
Electrostatic
discharges
(ESD)
e)
IEC 61000-4-2
±8 kV on contact
±15 kV in air
relative humidity should be at
Electrical fast
transients /
bursts
b) f)
IEC 61000-4-4
±1 kV on contact
frequency
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Surges Line-
to-ground
a)
IEC 61000-4-5 ± 2kV
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Conducted
disturbances
induced by RF
b) d) g)
IEC 61000-4-6
3 V
h)
6 V
h)
in the ISM bands
i)
c)
Portable and mobile RF
communication devices
should not be used near any
respect the recommended
equation applicable at the
frequency of the transmitter.
a) This test applies only to output lines intended to connect directly to outdoor cables.
b)
c)
d)
e)
position.
f) Capacitive coupling shall be used.
g)
range.
h)
i)