2.1 Unpacking / Operating Placement & Environment (continued)
Important
The following procedures must be followed exactly. Boule has no
responsibility in case of faulty or erroneous installation.
Installation/
Operating
Placement
The instrument should be placed in a laboratory environment according to the
guidelines below:
Place the instrument on a clean horizontal surface.
Avoid lifting the analyzer by the front cover.
Avoid exposure to sunlight.
Make sure the instrument has access to proper ventilation. The instrument
should have at least 2 inches (5 cm) of air above it.
Place the rear of the instrument so it has at least 4 inches (10 cm) of free
space behind it.
Figure 2.1
Installation/
Operating
Environment
Indoor Use
Temperature +64 to +90 ºF (+18 to +32 ºC)
Humidity < 80% Relative
Grounded main supply
Important
Operating the instrument in an environment over + 90°F (+32 °C) increases
service needs, as well as degradation of sample specimen.
4 inches
2 inches