EasyManua.ls Logo

Medonic M16C-US - Calibration

Medonic M16C-US
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
7.2 Calibration
Input Calibrator
Assay Values
Follow the instruction in Section 6.1 Quality Control to access the QC menu
and to input Control/Calibrator Assay Values from the Assay sheet.
Whole Blood
Calibration
The following instructions calibrate Open Tube, Cap Piercer, and Sampling
Device modes. Follow the instructions below to calibrate:
Step
Action
1
Prior to calibration, prime instrument by aspirating 2 fresh blood
samples. Disregard results.
2
Follow directions on Assay Sheet to scan in calibrator assay values.
3
Choose either List, Sample, or Main menu to begin calibrator analysis.
4
Using installed barcode reader, scan the Calibrator ID from the
calibrator vial label.
5
To perform calibration, it is recommended that five calibration analyses
be performed in consecutive order through the open tube mode.
Note
DO NOT use Cap Piercer or Autoloader mode to aspirate calibrator.
6
When analyses are complete press [ADVANCED] from the MENU tab.
7
Press [CALIBRATION] and then choose [WHOLE BLOOD].
Figure 7.1 Figure 7.2
Note
Calibration analysis must be last analysis performed on instrument for
parameter values to be shown in calibration menus. (e.g. no values will
show if in the middle of calibration a patient sample analysis was
performed)
Continued on next page
Important

Table of Contents