2.6 Power Supply
Warning
Electrical shock hazard.
Installation of external electrical equipment such as CVT must only be
carried out by authorized service engineers. Violating this might result in
injuries and/or loss of life and/or erroneous parameter results.
In case of
Symptom
Solution
High transient
voltage above
15%
-High background counts
on RBC, PLT or WBC.
-Defective instrument.
A CVT (magnetic stabilizer)
should be implemented to keep the
instrument from being damaged.
(If using an UPS, then it must have
surge protection.)
Guidelines
Guidelines are given in the Service Manual, “Installation auxiliary devices”
section. Contact your authorized distributor in such a case.
Power
interruptions
In case of an abrupt power loss there will be no damage done to the
instrument. Calibration constants and other parameters necessary for operation
are protected against main supply loss.
Before
connecting
In order to run the instrument, the frequency and main voltage needs to
correspond to user’s power outlet.
Locate the serial number plate on the rear of analyzer and check that the
main voltage and frequency corresponds to local main outlet.
If voltage and/or frequency does not correspond, then contact your
authorized distributor
Connecting
Power Adapter
Insert power adapter into the instrument’s main power inlet and connect it to
the main power supply. (This should only be performed after connecting the
reagent containers.)