12.3 Warning Displays (continued)
Assay Value Input failed. The Assay
sheet or order of scanning in the
barcodes may have been incorrect.
Verify that setups on the instrument
match the required setup for the
barcode reader. (See Section 4.3
and 6.1 for more detail.)
The barcode scanned in is not
recognized as a control sample in the
system. Verify that control sample is
being scanned in. (See Section 6.1
for more detail.)
12.4 Aspiration Issues
Description
This section contains information regarding errors associated with aspiration
and the aspiration probe.
If
Then
Possible cause
No aspiration of
sample is taking place.
1. Verify that there are no leaks and tubing
is connected properly and not kinked.
2. Perform valve check in Service Menu.
3. Perform clot prevention. See Section 8.2.
4. If clot prevention cycle does not work
contact technical service representative
for Manual Clot Removal procedure.
1. Blockage of tubing or leak
causes sample to not be pulled
correctly through shear valve.
2. Valve malfunction.
3. Clot in sample caused by
incorrect sample handling or
pathologic sample.
No cleaning of
aspiration probe
1. Suggest cleaning upper area of aspiration
probe.
2. Verify that there are no leaks and tubing
is connected properly and not kinked
1. Sample tube is touching the
upper part of the aspiration
probe when analyzing.
2. Diluent is not flowing correctly
through tubing to aspiration
probe.