EasyManua.ls Logo

Medonic M16C-US - Section 7: Calibration; Preparations before calibration

Medonic M16C-US
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
6.3 Initialization and Use of Xb Function
Description
The Xb function in the Medonic M-Series follows strictly the Bull algorithm
for the parameters MCV, MCH and MCHC. These parameters should not drift
as a function of time within a large patient population. The recommended range
setting is ± 3% from the expected mean value of these parameters.
Step
Action
1
Enter the QC menu and press [VIEW Xb STATS].
2
Select Xb points by Date or by default all sample data is selected.
3
Press [LJ VIEW] to display Xb L J diagrams.
Xb L-J
Diagrams
The image below is constructed from several samples and will not be
shown as below until a sufficient amount of samples have been analyzed.
Figure 6.9 Figure 6.10
4
Select [MORE] to view selected conditions and matched ranges.
5
Print diagrams by choosing [PRINT].
6
To change ranges on Xb Diagrams go to Setup Menu 3 and press [XB
RANGE SETUP]. Here operator can change low and high ranges on the
three parameters. To update or change Xb range setup input the
Authorization Code [2576].
Reference
Bull BS, Hay KL. The blood count, its quality control and related methods: X-bar calibration and
control of the multichannel hematology analysers. In: Clangoring I. editor. Laboratory
Hematology: An account of Laboratory Techniques. Edinburgh.

Table of Contents