EasyManua.ls Logo

Medonic M16C-US - Power Supply

Medonic M16C-US
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
2.4 Reagent Installation (continued)
Step
Action
1
Select MENU tab.
2
Press [REAGENT SETUP] and then press [ENTER NEW REAGENTS].
3
Scan in barcodes on reagent containers, when all barcodes are entered a
screen will display that reagent barcodes have been accepted.
Figure 2.11 Figure 2.12
4
Return to MAIN Menu and press [ADVANCED].
5
Press [MAINTENANCE] and then [FILL SYSTEM].
Figure 2.13 Figure 2.14 Figure 2.15
6
The system is now filling up with reagents. This cycle will last for
approximately 3 minutes.
Print All
Settings
After initial setup, it is recommended to print all analyzer settings and keep for
personal records. Select [ADVANCED] from Main Menu, then [SETUP], and
then [PRINT ALL SETTINGS].
Factory
Calibration
All sample analysis modes (open tube, pre-dilute, MPA, cap piercer, sampling
device) are factory calibrated. However, calibration should always be checked
upon installation. See Section 7 for more details.

Table of Contents