12.3 Warning Displays (continued)
Capillary Device Warning Displays
The MPA was opened during an
inappropriate time. It is
recommended to perform a prime
cycle before next analysis.
The MPA was opened during a cycle
or analysis. Re-insert holder, and
follow suggested recommendation.
The MPA holder was opened during
an inappropriate menu. The MPA
holder should only be opened in
List, Sample or Main menu.
Cap Piercer and Autoloader Warning Displays
The Cap Piercer door was opened
before the CAP door lock had been
activated. Close the Cap Piercer
door to continue with the analysis.
The aspiration wheel has been
interfered with during mixing. Press
[OK] to return to the sample menu. To
proceed with the analyses press
[CONTINUE] in Sampling Device
List Menu.
Three aspirations have been
attempted. All have failed. Verify
that sample tubes contain at least 1
ml of blood. (See Section 5.9 for
more detail.)