EasyManua.ls Logo

Megger DET2/3 - Instrument Calibration and Accuracy; Instrument Calibration Check Procedure; Checking Instrument Accuracy

Megger DET2/3
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DET2/3
31
Calibration check tools
Calibration check tools
The instrument's calibration should be checked, before and after each test, against the calibration check tool.
Instrument calibration check
1. Make sure that the mode switch is set to Off.
2. Connect the instrument as shown:
C1
CAT IV 300V
P1
P2
C2
E
ES
S
H
I
V
E
X
Y
V
MVC1010 MVC1010
MCC1010
I
Calibration
adaptor
3. Set the function switch to 2P, 3P or 4P.
4. Start a test:
Press and release TEST.
The instrument completes pre-measurement checks.
The resistance is shown and should match the value written on the calibration check tool.
Checking instrument accuracy
Instrument accuracy: 0.5% (+ 2 digits). At 25 Ω is this gives an allowed measurement range of +/- 0.145 Ω.
Calibration check tool accuracy: 0.1%. At 25 Ω this gives an allowed variation of 0.025 Ω.
Therefore a test, which uses the 4 pole position, will produce the following bands:
25 + 0.145 + 0.025 = 25.17 (a figure greater than this is out of specification)
25 + 0.145 – 0.025 = 25.12 (a figure between 25.12 and 25.17 is possibly out of specification)
25 – 0.145 + 0.025 = 24.88 (a figure between 24.88 and 25.12 is in specification)
25 – 0.145 – 0.025 = 24.83 (a figure less than this is out of specification)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Related product manuals