D33451 | Date 01/2018 13
Construction and description
2.7 Technical data
Appliance size 50 105 150 240
Appliance width D* [mm] 559 719 719 759
Appliance height E* [mm]
(varies due to adjustable feet)
791 846 1066 1176
Appliance depth F* (without door handle) [mm] 521 591 591 691
Depth of door handle [mm] 56
Chamber width A* [mm] 400 560 560 600
Chamber height B* [mm] 425 480 700 810
Interior depth C* [mm] (less 35 mm for fan) 330 400 400 500
Chamber volume [litres] 56 107 156 241
Net weight [kg] 55 75 90 110
Weight including packaging [kg] 74 100 116 145
Power [W] 1100 1300 1500 1650
Supply voltage [AC V] 230/115
Frequency [Hz] 50/60
max. number of sliding shelves 5 6 10 12
max. load per sliding shelf [kg] 15
max. load per appliance [kg] 75 90 120 140
Operating temperature range
5 °C above room temperature up to
50 °C
Setting temperature range [°C] +18 to +50
Adjustment precision [°C] 0.1
Temporal temperature deviation
(in accordance with DIN 12880:2007-05) at 37 °C
[K]
±0.1
Spatial temperature deviation at +37 °C
(in accordance with DIN 12880:2007-05) at 37 °C
[K]
±0.3
Adjustment range of active humidity control
[% rh]
(only for appliances with the corresponding con-
figuration)
40 to 97 and rh off
Setting accuracy humidity [%] 0.5
Adjustment range for CO
2
[%]
0 to 20
Setting accuracy for CO
2
[%]
0.1
Adjustment range for O
2
[%] (optional and only for
appliances with active humidity control)
1 to 20
Setting accuracy for O
2
[%]
0.1
* see Fig. 6 on page 14 .