48 D33451 | Date 01/2018
Menu mode
All other settings can be made accordingly. The settings possible are described in the follow-
ing sections.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati-
cally restores the former values.
7.3 Setup
7.3.1 Overview
In the SETUP display, you can set the following parameters:
the IP address and subnet mask of the appliance's Ethernet interface (for connection to a
network)
the temperature display unit (°C or °F, see page 49)
how the digital backwards counter with target time setting works (Timer Mode, see page
50)
Remote control (see page 50)
Gateway (see page 50)
If the Setup menu contains more entries than can be
displayed, this is indicated by the display "1/2". This
means that there is a second "page" of entries.
To display the hidden entries, use the turn control to scroll
beyond the lowest entry. The page display changes to "2/2".
7.3.2 IP address and subnet mask
If you want to operate one ore more appliances in a network, each appliance must have its
own unique IP address for identification. By default, each appliance is delivered with the IP
address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
1
9
2.1
68
.
1
Editor
Programmname
Simulation Protokoll
-
+
-
+
INP 250 Test 01
180
.0°C
i
HPP 250 Labor
i
37
.0°C
44
.4%rh
15
15
STAND BY
Programm negnulletsniEllokotorPtäreG efliHnekcurD
AtmoCONTROL
Fig. 26 Operation of several appliances in a network (schematic example)
Setup
Unit
IP adress 255.145.136.22
5
Subnet mask 255.255.0.0