EasyManua.ls Logo

Memmert ICO - Putting into operation; Connect to power supply

Memmert ICO
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 D33451 | Date 01/2018
Putting into operation
4. Putting into operation
NOTICE
The first time the appliance is operated, it must not be left unattended until it has reached the
steady state.
4.1 Connect the appliance to the power supply
WARNING
Condensation might cause a short circuit. After transport or storage
at high humidity conditions, the appliance shall be stored unpacked
at normal conditions for at least 24 hours. During this period of time
the appliance shall not be connected to the power supply.
Observe the country-specific regulations when making
connections (e.g. in Germany: DIN VDE 0100 with earth
leakage circuit breaker). Observe the connection and
power ratings (see nameplate and "Technical Data" on
page 13). Make sure to establish a safe PE conductor
connection.
Plug the provided power cable into the rear of the appli-
ance and connect it to the power supply ( Fig. 9 ). Place the
power cable so that
it is easily accessible at all times and can be pulled off quickly, for example in case of inter-
ference or an emergency;
it does not represent a trip hazard;
it cannot come into contact with any hot parts.
4.2 Establishing water supply
4.2.1 Water speci cations
For humidification with a water tray or steam generator, only use: demineralised water (aqua
dem) in accordance with VDE 0510/DIN EN 50272 if you are working according to these
regulations (production conductivity 5 10 μS/cm). You can buy battery water in accord-
ance with the VDE 0510 standard in all larger drug stores, supermarkets and hardware stores
as well as from wholesale traders. The VDE 0510/DIN EN 50272 standard has to be explicitly
specified on the label.
Otherwise, possible calcification in the steam generators, steam piping and steam generator
pumps could affect the correct functioning of the appliance.
Only use chlorine-free water with a pH value of > 5 and < 7.
4.2.2 For appliances with passive humidity control: inserting the water tray
Mount the provided sealing lip to the narrow side of the water tray (Fig. 10) and fill 1.5 cm to
2 cm of water into the tray (for specifications see section 4.2.1).
Place the water tray with the attached sealing lip in the centre of the appliance floor and
carefully push it towards the rear panel until the sealing lip is completely under the ventilation
opening, touching the rear panel along its entire length (Fig. 11). The sealing lip magnetically
adheres to the rear panel and returns the water condensing on the humidity limiter to the
water tray.