44 D33451 | Date 01/2018
Malfunctions, warning and error messages
Description Cause Action See
Displays suddenly look
different
Appliance is in "wrong"
mode
Change to operating
or menu mode by
pressing the
MENU
key
Error message
T:E-3 in
the temperature display
TEMP
T:E-3 Set 37.0 °C
37.4°C
Temperature operating sen-
sor is defective. The moni-
toring sensor takes over the
measurement function.
► The appliance can
temporarily be
kept in service
► Contact customer
service as soon as
possible
Page 2
Error message
AI E-3 in
the temperature display
TEMP
AI E-3 Set 37.0 °C
37.4°C
Temperature monitoring
sensor is defective. The
operating sensor takes over
the measurement function.
► The appliance can
temporarily be
kept in service
► Contact customer
service as soon as
possible
Page 2
Error message
E-3 in
the temperature display
TEMP
Set 45.0 °C
E-3 °C
Operating and monitoring
sensor defective
► Switch off appli-
ance
► Remove the cham-
ber load
► Contact customer
service
Page 2
Error message
E-6 in
the humidity display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective ► No humidity con-
trol possible
► Contact customer
service
Page 2