Menu mode
Example: Temperature deviation at 42 °C should be corrected.
1. Press the activation key to the right of the
CALIB display. The display is enlarged
and the temperature adjustment option
is automatically selected.
Cal1 °C -0,2K
Cal2 100.0 °C +0,1K
Cal3 180.0 °C -0,2 K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
CK0,2-
CK0,1+
CK0,2-
Calibrate
Cal1
Cal2
Cal3
40.0
50.0
30.0
Temperature
Humidity
CO2
O2
2. Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
CK0,2-
CK0,1+
CK0,2-
Cal1
Cal2
Cal3
50.0
30.0
40.0
3. With the turn control, set the calibration
temperature Cal2 to 42 °C.
CK0,2-
CK0,1+
CK0,2-
Cal1
Cal2
Cal3
50.0
30.0
42.0
4. Save the setting by pressing the confir-
mation key. The corresponding calibra-
tion value is automatically highlighted.
CK0,2-
CK
CK
0,2-
Cal1
Cal2
Cal3
50.0
30.0
42.0 0,1+
5. Set the calibration value to 0.0 K and ac-
cept the setting by pressing the confirma-
tion key.
CK0,2-
CK
CK
0,2-
Cal1
Cal2
Cal3
50.0
30.0
42.0 0,0
6. Position the sensor of a calibrated refer-
ence instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode,
adjust the set temperature to 42 °C.
21.4°C
TEMP
42.0 °CSet
8. Wait until the appliance reaches the set
temperature and displays 42 °C. The ref-
erence instrument should display 43.6 °C.
42.0°C
TEMP
Set 42
.0°C
43.6 °C